Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The EasyStand Bantam Medium is intended for use in the electromagnetic environment specified below. The customer
or the user of the EasyStand Bantam Medium should assure that it is used in such an environment.
Emissions Test
RF emissions CISPR 11
Group 1 The EasyStand Bantam Medium uses RF energy only
for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interfer-
ence in nearby electronic equipment.
The EasyStand Bantam Medium is suitable for use
in all establishments, including domestic establish-
ments and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
Compliance Electromagnetic Environment - Guidance
RF emissions CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The EasyStand Bantam Medium is intended for use in the electromagnetic environment specified below. The customer
or the user of the EasyStand Bantam Medium should assure that it is used in such an environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete, or ceramic tile. If
floors are covered with synthetic material, the rela-
tive humidity should be at least 30 %.
Mains power quality should be that of a typical com-
mercial or hospital environment.
IEC 60601
Test Level
Electromagnetic Environment - Guidance
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output
lines
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Compliance
Level
± 2 kV for power
supply lines
± 1 kV for input/
output
lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the EasyStand Bantam Medium requires continued
operation during power mains interruptions, it is
recommended that the EasyStand Bantam Medium
be powered from an uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Pow’r Up Option
Page 27